OPERATING INSTRUCTION MANUAL
MODEL 2200C
CONDUCTIVITY ANALYZER
N116-9 REV. 8.0
Water Analytics, Inc.
100 School Street
Andover, MA 01810
(978) 749-9949
(855) 747-7623
Fax (978) 749-9961
www.Wateranalytics.net
Vue de la page 0
1 2 3 4 5 6 ... 27 28

Résumé du contenu

Page 1 - N116-9 REV. 8.0

OPERATING INSTRUCTION MANUAL MODEL 2200C CONDUCTIVITY ANALYZER N116-9 REV. 8.0 Water Analytics, Inc. 100 School Street Andover,

Page 2 - TABLE OF CONTENTS

4.5.1 The analog output signals consist of a non-isolated 0-1 mA, 0-5 Vdc, and isolated 4-20 mA signals. 4.5.2 From th

Page 3 - INSTRUCTION MANUAL

4.7.1 The controls which are frequently used in the normal operation of the instrument are all accessible on the control panel. Some switches, whic

Page 4

4.8.1 Output hold, is a function which freezes all output signals at the last value to prevent the occurrence of wild distortions during programming

Page 5 - 3.0 INSTALLIATION

5.1.1 To enter the menu press CALL and the PASSWORD LED will illuminate. If the password feature has been disabled by the dip switch, TEST will be t

Page 6

describe this setup is by example. Suppose you wish the outputs to span from 400uS to 600uS. Proceed as follows: a) Enter the menu by

Page 7

increasing conductivity, the deadband will be lower than the setpoint. i.e. the auxiliary device connected to the relay will decrease the conductivit

Page 8

a) Press CALL to enter the menu, proceed by pressing CALL until the ALARM HIGH LED is illuminated. Use the arrow keys make the display read the desi

Page 9

main circuit board, in the up position, (this places a 3000 ohm across the RD and GN terminals) the temperature should read 25°C or 77°F.

Page 10

6.1.2 Be sure the cell cable is well protected. The cell cable should run in conduit but never in the same conduit with line power. Excess cable s

Page 11

• Adjust the offset and span of the 4-20 mA output 7.2 Access to Utility Menu 7.2.1 The Utility Menu is prot

Page 12

TABLE OF CONTENTS SECTION DESCRIPTION PAGE 1 GENERAL INFORMATION 1 2 SPECIFICATIONS 2 3 INSTALLATION 3 4 DESCRIPTI

Page 13

0-20 uS 0.05 0-50 uS 0.05 0-100 uS 0.05 0-200 uS 0.5 0-500 uS 0.5 0-1000 uS 0.5 0-2000 uS 1.0 0-5000 uS 1.0 Range mS Required Cell Constant 0-10 mS

Page 14

b) Press CALL until the OUTPUT HIGH LED illuminates, which is “Temperature output, 100% point” in the Utility Menu (See table in 7.2.2). Use the ar

Page 15

a) Turn off the power to the instrument. Connect your DVM in series with the 4-20 mA output terminals on the power supply board. b) Turn on the p

Page 16

8.2 Invalid Calibration 8.2.1 Invalid calibration will be indicated when the input is out of range. This could occur i

Page 17 -

9.1.2 If your 2200C appears dead or intermittent, check the breaker, make sure that the instrument is set up for the available line voltage and make

Page 18

Range uS Required Cell Constant R68 Value 0-2uS 0.01 10,000 0-5 uS 0.01 4000 0-10 uS 0.05 10000 0-20 uS 0.05 5000 0-50 uS 0.05 2000 0-100 uS 0.05 10

Page 19

b) The "ESCAPE" procedure is to be used if a unit is serviced for a new processor, or in case of a memory loss problem, when so advised b

Page 20

c) The microprocessor circuit board is located on the swing-out assembly behind the door. Swing the assembly out, lift it up to unseat the lower

Page 21

computing device pursuant to Subpart J of Part 15 of FCC Rules, which are designed to provide reasonable protection against such interference when op

Page 22 - 8.0 DIAGNOSTICS

Model 2200C Page 1 MODEL 2200C CONDUCTIVITY CONTROLLER INSTRUCTION MANUAL 1.0 GENERAL INFORMATION

Page 23

2.0 SPECIFICATIONS DISPLAY: 4 digit LED, 1/2" high digits MEASURING RANGES: Eighteen ranges; The range is set

Page 24

3.0 INSTALLIATION 3.1 Location 3.1.1 Locate the instrument within 3000 feet of where the AquaMetrix

Page 25

3.4 Electrical Connections CAUTION: The instrument operates from line voltage. This constitutes a possible shock hazard.

Page 26

3.5 Cell Connections 3.5.1 Connect the 4 wires of the cell cable to the TB2 terminal strip on the main circuit board b

Page 27

4.0 Description of Functions 4.1 Overview 4.1.1 The Model 2200C is a microprocessor based conductivity

Page 28

4.2 Calibration 4.2.1 All conductivity systems need to be calibrated when first placed in service and thereafter from